  1. questionCan the gas manifold be used with the horizontal diffusion system?
    The gas manifold is not normally used with the horizontal chamber. Normally, if a gassed solution needs to be introduced into the horizontal chamber it will first be bubbled into the media, which is then pumped into the basolateral chamber. Since the gas manifold is basically twelve needle valve ...

  2. questionHow do I know what Ussing System components and accessories I will need?
    The Ussing System brochure does a good job of making sure you don't miss anything as you put a system together. Please see the PDF.

  3. questionThe bridge fittings for my ussing chamber are too long
    The bridge fittings are deliberately cut long to allow you to precisely size them for your setup (excised membrane or culture cup). Simply trim the tip to the correct size, taking into account that the voltage electrode needs to be as close to the membrane as is feasible.Also, note that the meta ...