  1. questionHow do you insert the cell culture dish adapter rings into the DH-35i, DH-40i, QE-1 and QE-2 microincubators?
    The cell culture dish adapter rings are tapered for a precise fit and should be mounted with the notch facing down. If the ring is not oriented correctly it will be very difficult to insert and remove.A special tool is provided to assist in removing the adapter ring from the microincubator platf ...

  2. questionWhich color adapter rings go with which cell culture dish when using the DH-35i, DH-40i, QE-1 and QE-2 micrincubators?
    Color codes for the different adapter rings are shown below. Color ID OD Gold 1.534 1.599 Green 1.415 Blue 1.331 Red 1.474 Black 1.396 Silver 1.394

  3. questionCan a CC-28 thermistor be repaired?
    Yes, both the copper-colored monitor and white feedback thermistors can be replaced. The replacement parts come with instructions for attaching to the CC-28 cable. TS-60P 64-0269 White feedback thermistor TS-70B 64-0270 Copper-colored monitor thermistor

  4. questionDon't see your answer? Please submit a question!
    We welcome your questions! Submit a question to us (via email) by clicking here and we'll answer you directly. We'll also anonymously post your question to our FAQ Knowledge Base (KB) to assist other people.If you choose to not communicate via this KB, then please send your questions directly to ...