Can Warner Series 20 Perfusion Chambers be used with #0 thickness Cover Glass
Nothing speaks against the use of #0 thickness glass coverslips with our Series 20 chambers. However, you might want to use a P- or PH- platform with screw-clamps rather than a PM- platform with magnetic clamps. The screw- clamps will allow you to adjust the clamp pressure. Be careful not to br ...
Is it possible to overheat a heated platform?
Heated platforms use a feedback thermistor (the big white one on the CC-28 cable) that is placed into the platform body. With this thermistor in place the controller will not allow the platform to overheat.In any case, heating a platform beyond what the controller will allow will not cause any a ...
What chambers do the Sutter chamber columns accommodate?
The Sutter MT-150/W2 chamber column accommodates any Series 20 chamber and platform.You must inform Sutter if you intend to use a Warner PM series platform as they will supply a different adapter.The MT-150/CUST is a custom built Sutter chamber column that is designed to accommodate a user speci ...
Can a CC-28 thermistor be repaired?
Yes, both the copper-colored monitor and white feedback thermistors can be replaced. The replacement parts come with instructions for attaching to the CC-28 cable. TS-60P 64-0269 White feedback thermistor TS-70B 64-0270 Copper-colored monitor thermistor
Can a non-heating Series 20 Platform be converted into one that heats?
The RH-2 heater elements (part number 640274) can be used to convert almost any series 20 platform into one that can be heated. The elements simply screw into the platform.However, some of the older platforms cannot accommodate the elements because there are no holes for screws. Check to make su ...
Can a platform be heated without using the feedback thermistor?
Yes, however, the feedback thermistor is needed for automatic temperature regulation. If it is omitted, then only manual temperature control is available.The white feedback thermistor from the CC-28 cable is inserted into the platform body to allow automatic temperature control. Without the whit ...